Hamilton Medical Recalls Coaxial Breathing Circuit Set | RT (opens in a new tab)
Hamilton Medical is recalling certain lots of coaxial breathing sets used to connect Hamilton ventilators to a patient tube or respiratory mask during ventilation, according to an FDA alert. The FDA says Hamilton Medical Inc sent all affected customers an Urgent Medical Device Recall notice for certain lots of coaxial breathing sets due to a quality issue involving a defective blade in the tube-cutting stage of production that may lead to a crack in the inner blue limb of the circuit. According to the FDA, this inner blue limb delivers fresh gas to the patient while the outer limb carries away exhaled gas and leakage from a cracked blue limb can lead to partial or complete rebreathing of exhaled gases. Hamilton Medical ventilator coaxial breathing circuits are intended to connect Hamilton-C1/T1/MR1 ventilators to a patient tube or respiratory mask during ventilation, the FDA says.